Clinical Chemistry Automatic PKL PPC 200

24 hour non-stop cooling system to ensure reagent at 2-8¡
Durable ceramic syringes to ensure accuracy & precision
High accurate optical system
Collision protection in both vertical and horizontal directions, stop & alarm automatically once touching barrier, not affect former tests
Automatic eligible cuvettes detection & selection
Select best test point by reaction curve, create new factor automatically
Support LIS interface

depliant PKLPPC200

Description

Assay methods: End Point, Kinetic, Fix time etc.

Principle: Photoelectric colorimetry

Light Source: Halogen lamp 12V/20W

Photometry range: 0~3.2Abs

Resolution: 0.0001

Abs Wavelength: 340nm\405nm\450nm\492nm\510nm\546nm\578nm\630nm\700nm\800nm

Throughput: 200 tests/hour

Reagent tray: 60 reagent positions (include1detergent position,1 dilution position)

Sample tray: 45sample positions, including detergent, standard, QC, STAT positions

Reaction tray: 120 reaction cuvettes Sample volume: 2~100ul, with 0.1ul increment

Reagent volume: R1:10~500uL, R2: 10~500uL,with 0.5uL increment Minimum
reaction volume: 180uL

Maximum reaction time: 10 minutes

Water Consumption: 5L/hour under working status

Clean unit: 8-step auto-washing system with detergent

Calibration: Calibration reset, select best test point by reaction curve, no need second calibration Line/non-line;multi-standards assay

Control rules: Westgard multi-rule, Cumulative sum check, Twin plot 3 level controls
for each ifem, analyxing and printing QC analysis diagram

Temperature control: incubator 37¡0.1¡ Power supply: AC 230(1¡10%)V, 50/60Hz, 500VA

Ambient: Operating temperature: 10~30¡

Relative humidity: ¡85% Atmospheric pressure: 86~106kPa